请选择 型号/数量
INPUT |
||
Voltage |
115/230 VAC ± 10%, User selection |
|
Frequency |
50/60 Hz ± 5% |
|
Fuse |
6.3 Amp 250V Slo-Blo |
|
DIELECTRIC WITHSTAND TEST MODE |
||
Output Rating |
5 KV @ 20mA AC |
|
6 KV @ 5 mA DC |
||
Output Adjustment |
Range: |
0-5 KV AC, 0-6 KV DC |
Resolution: |
10 volt/step |
|
Accuracy: |
± (2% of setting + 5 volts) |
|
Voltage Display |
Range: |
0 - 5 KVAC / 0-6KV DC |
Resolution: |
10 volt/step |
|
Accuracy: |
± (2% of reading + 10 volts) |
|
HI / LO -Limit AC |
Range: |
0.00 - 20.00 mA |
Resolution: |
0.01 mA/step |
|
Accuracy: |
± (2% of setting + 0.02 mA) |
|
AC Current Display |
Range: |
0.00 - 20.00 mA |
Resolution: |
0.01 mA/step |
|
Accuracy: |
± (2% of reading + 0.02 mA) |
|
HI / LO - Limit DC |
Range: |
0.00 - 5.00 mA |
Resolution: |
0.01 mA/step |
|
Accuracy: |
± (2% of setting + 0.02 mA) |
|
DC Current Display |
Range: |
0.00 - 5.00 mA |
Resolution: |
0.01 mA/step |
|
Accuracy: |
± (2% of reading + 0.02 mA) |
|
Failure Detector |
Audible and Visual |
|
Output Ripple |
< 5% (6KVDC/5 mA at Resistive Load) |
|
Discharge Time |
< 200 ms |
|
Maximum Capacitive Load in DC Mode |
1µF < 1KV , 0.75µF < 2KV, 0.5µF < 3KV, 0.08µF < 4KV, 0.04µF < 5KV, 0.01µF < 6KV |
|
AC Wave Form |
Sine wave Distortion : < 2% THD ; Crest Factor= 1.3 – 1.5 |
|
Output Frequency |
Range: |
60 or 50 Hz, User Selection |
Output Regulation |
± (1 % of output + 5 volts) from no load to full load |
|
Dwell Timer |
Range: |
0, 0.2 - 999.9 sec, (0 = Constant) |
Resolution: |
0.1 sec increments |
|
Accuracy: |
± (0.1% + 0.05 sec) |
|
Ramp Timer |
Range: |
0.1 - 999.9 sec |
Resolution: |
0.1 sec increments |
|
Accuracy |
± (0.1% + 0.05 sec) |
|
INSULATION RESISTANCE TEST MODE |
||
Output Voltage |
Range: |
100 - 1000 Volts DC |
Resolution: |
1 volt/step |
|
Accuracy: |
± (2% of reading + 5 volts) |
|
Voltage Display |
Range: |
0 - 1000 V |
Resolution: |
1 volt/step |
|
Accuracy: |
± (2% of reading + 2 counts) |
|
Resistance Display |
Range: |
1 - 1000 MΩ (4 Digit, Auto Ranging) |
Resolution: |
0.01/0.1 MΩ |
|
Accuracy: |
± (3% of reading + 2 counts) at test voltage > 500V |
|
± (7% of reading +2 counts) at test voltage < 500V |
||
HI-Limit |
Range: |
0, 1 - 1000 M Ω (0 = Off) |
LO-Limit |
Range: |
1 - 1000 M Ω |
Delay Timer |
Range: |
0, 0.5 - 999.9 sec (0 = Constant) |
Resolution: |
0.1 sec/step |
|
Accuracy: |
± (0.1% + 0.05 sec) |
|
GROUND BOND TEST MODE |
||
Output Voltage |
Range: |
AC 6 volts Max. |
Output Frequency |
Range: |
60 or 50 Hz, User Selection |
Accuracy: |
±0.1% |
|
Output Current |
Range: |
3.0 - 30.0 Amps AC |
Resolution: |
0.1 Amp/step |
|
Accuracy : |
± (2 % of setting + 0.1A) |
|
Current Display |
Range: |
0.0 - 30.0 Amps |
Resolution: |
0.1 Amp/step |
|
Accuracy: |
± (3 % of reading + 0.1A) |
|
Resistance HI-Limit |
Range: |
0 - 510 mΩ |
Resistance Display |
Range: |
0 – 510 mΩ for 0.0 – 10.0 Amps |
0 – 200 mΩ for 10.1 – 25.0 Amps |
||
0 – 150 mΩ for 25.1 – 30.0 Amps |
||
Resolution: |
1 mΩ/step |
|
Accuracy: |
± (2 % of Setting + 2 mΩ ) |
|
Dwell Timer |
Range: |
0, 0.5 - 999.9 sec (0 = Constant) |
Resolution: |
0.1 sec/step |
|
Accuracy: |
± (0.1% + 0.05 sec) |
|
Lead Resistance Offset |
Max. Offset Capability: |
100 mΩ |
Resolution: |
1 mΩ / step |
|
Accuracy: |
± (2 % of Setting + 2 mΩ ) |